3.4 Checking the motors
Tab 3
3.4
Checking the motors
Check operating status of motors.
Visual inspection:
• Is the motor intact?
• Are the plug connections OK?
Temporarily install replacement motor
and check unit function.
• Is the unit functioning properly?
Replace motor, see section 6.2 or
section 6.3.
Replace the motor and/or fix the plug
connections.
no
Motor is OK!
Continue troubleshooting according
to error list (see section 2.5).
no
yes
yes