Tab 3 3.8 Error analysis of X-RAY control signal path: up to unit hardware revision AG
3.8
3.8.3
Error messages
with and without installed
remote control
Error code
Description
Actions required
see page
E3
41
20
Faulty detection of release signal by
DX41 processor when the unit is
switched ON.
z
Replace board DX41
E6
41
23
Faulty detection of release signal by
DX41 processor during operation of
the unit.
z
Replace board DX41
E6
41
25
The DX41 detects no release signal
when the exposure is started.
z
Replace board DX41
E3
13
40
Short circuit in signal path between
boards DX11 and DX41 during
switch-ON.
z
Replace cable L17
z
Replace board DX41
z
Replace board DX1
z
Replace board DX11
E6
13
41
Release signal missing on
board DX11 at start of exposure.
z
Replace cable L17
z
Replace board DX41
z
Replace board DX1
z
Replace board DX11
E3
13
42
Short circuit in signal path between
boards DX11 and DX41 during opera-
tion of the unit.
z
Replace cable L17
z
Replace board DX41
z
Replace board DX1
z
Replace board DX11
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...