59 38 399 D3352
4 – 12
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
4.3.3
Adjusting the pan sensor
(XG
Plus
/ XG 5 / 3 / 3 PPE)
Menu: Diaphragm/system adjustment
‡
Pan - Sensor adjustment
Preparations
z
Plug the sensor into the sensor slot on the panoramic X-ray unit.
z
Insert the needle phantom in the bite block holder of the panoramic X-ray
unit (see page 4-9).
ORTHOPHOS XG 3 / 3 PPE:
z
The fixed diaphragm must be dismantled prior to the sensor adjustment
for the ORTHOPHOS XG 3 / 3 PPE system version (see page 4-10).
Opening
PAN
-
SENSOR
ADJUSTMENT
1.
Open the
PAN
-
SENSOR
ADJUSTMENT
submenu (see section 4.2).
NOTE
i
The menu provides a precision adjustment and a coarse adjustment
(precision adjustment is preset). Perform a precision adjustment first. In most
cases, previous coarse adjustment is not necessary.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...