59 38 399 D3352
4 – 20
D3352.076.01.13.02
07.2008
4.3 Adjusting the panoramic X-ray unit
Tabs 4
Manual adjustment of the PAN sensor
The manual adjustment procedure is similar to the one for automatic adjust-
ment. The only difference is that the default adjustment values automatically
determined by SIDEXIS are overwritten by manually determined adjustment
values in the
PAN
-
SENSOR
ADJUSTMENT
submenu.
Determining the adjustment values for
S1, S2 and S3 manually
z
Measure distances
S1
,
S2
and
S3
with the SIDEXIS measuring ruler.
NOTE
i
To measure S1, S2 and S3, estimate the horizontal center position of the dis-
played needles. Measure in the lower area of the needles if possible, since
they may have been bent slightly after repeated use.
Tip:
To facilitate the measuring procedure, you can color the image in SIDEX-
IS (see also SIDEXIS Operator’s Manual).
S1
S2
S3
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...