59 38 399 D3352
4 – 68
D3352.076.01.13.02
07.2008
4.4 Adjusting the cephalometer (XG
Plus
/ XG 5, if ceph is installed) Tabs 4
4.4.4
Adjusting the ceph fixed point of rotation
Menu: Diaphragm/system adjustment
‡
Ceph - Fixed point of rotation
CAUTION
It is essential that the PAN needle phantom is removed from bite block holder
of the panoramic X-ray unit before a Ceph exposure is taken; otherwise the
phantom may collide with the sensor.
Preparations
z
Remove the ceph test phantom from the sensor slot of the panoramic X-
ray unit (see page 4-57).
Opening
CEPH
-
FIXED
POINT
OF
ROTATION
1.
Go to the
CEPH
-
FIXED
POINT
OF
ROTATION
submenu.
NOTE
i
The menu provides a precision adjustment and a coarse adjustment
(precision adjustment is preset). Perform a precision adjustment first. In most
cases, previous coarse adjustment is not necessary.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...