Manual adjustment of the TSA sensor
The manual adjustment procedure is similar to the one for automatic adjust-
ment. The only difference is that the default adjustment values automatically
determined by SIDEXIS are overwritten by manually determined adjustment
values in the
TSA
-
SENSOR
submenu.
Determining adjustment values for Sx
and Sy manually
z
Measure distances
Sx
and
Sy
with the SIDEXIS measuring ruler.
NOTE
i
Measure from the edge of the imaged diaphragm area roughly up to the edge
of the image area. Leave a small border, since a surrounding white border
should be present following the correction.
S2
Sy
Sx