5.5 Service routine S002
Tabs 5
S002.3: Radiation w/o rot. movement, max. radiation time selectable, prim. diaphr. opened fully
5.5.2
S002: Test step 3
(not for XG 3 / 3 PPE)
Difference from S002, test step 1:
9
Primary diaphragm is opened fully
NOTE
i
For selection and operating sequence, see S002, test step 1.
After you have selected test step 3, the primary diaphragm is moved to
the new position (opened fully).
During this time, a progress indicator is displayed in selection field 1.