5 – 33
Tab 5
5.6 Service routine S005
5.
6
S005.5: Temperature sensor test
5.6.3
S005: Test step 5
Selecting test step 5
1.
Select test step 5 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
After test step 5 is selected, selection field 1 displays the single tank
temperature in
°
C. The display is updated once per second.
Quitting the service routine
XG
Plus
:
To return to the service routine selection menu, press the
Service key
or
the
double arrow key
.
To quit the Service menu and return to the main menu, press the
double arrow key
.
For XG 5 / 3 / 3 PPE:
To return to the service routine selection menu, press the
up arrow key
above selection field 3.
To quit the Service menu and return to the Main menu, press the
up arrow key
above selection field 3.
1
2
3
+
-
S005
5
n
(
1
2
3
S005_______5____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
+
-
(
n
(
n
(
o
(
o
1
2
3
31.5C
n
(
1
2
3
__31.5_C________
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
S5
5
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...