6.8 Replacing/adjusting the MS light localizer laser module (PAN)
Tabs 6
6.8
Replacing/adjusting the MS light localizer
laser module (PAN)
Removing the covers
z
Remove the covers (see also section 1.14):
– Top arm cover
z
Pull the cables of the laser modules (FH+MS) off of connectors
X807
(FH)
and
X811
(MS) on board
DX1
.
z
Detach the cover (see also section 1.14):
– Front slide cover, carefully pulling it together with the FH light localizer
off toward the front.
Removing the defective laser module
z
Detach the cable from the cable holder.
1.
Loosen the two screws
A
and carefully remove the laser module (MS) to-
ward the front.
Installing the new laser module
z
Insert the new laser module (MS) in the holder and tighten the screws
A
loosely.
NOTE
i
Make sure that the lines in the laser module run horizontally.
z
Plug the cable of the new MS laser module into connector
X811
on board
DX1
and run the cable in the cable holder.
NOTE
i
Before reattaching the front slide cover, you must adjust the MS light localizer.
DX1
A
1.
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...