59 38 399 D3352
6 – 90
D3352.076.01.13.02
07.2008
6.20 Replacing circuit boards
Tabs 6
6.20.4
Check jumper on circuit board DX42
(replacement of board DX42)
From system hardware version BA (November 2006), the DX42 board will
have two jumpers which are configured through the system hardware version
with
or
without
board DX41.
If board DX42 is supplied as a spare part (from November 2006), the jumpers
will be set to operation
without circuit board DX41
. If a new DX42 circuit board
(supplied during/after November 2006) is to be inserted in a system with a
system hardware version <BA as a spare part, the jumpers must be reconfig-
ured prior to installation in accordance with the following
steps.
6.20.5
Jumper position on circuit board DX42
Unit operation
without DX41
(factory setting from November 2006,
hardware system version
≥
BA)
Unit operation
with DX41
(factory setting until October 2006,
system HW version
<
BA)
DX42
DX41
IN
DX
41
I
N
X110
X109
Jumper outside
DX41 IN
DX41 IN
X110
X109
Jumper inside
Summary of Contents for ORTHOPHOS XG 3 DS
Page 4: ......
Page 9: ...ORTHOPHOS XG 1General information...
Page 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Page 59: ...ORTHOPHOS XG 2 Messages...
Page 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Page 125: ...ORTHOPHOS XG 3 Troubleshooting...
Page 153: ...ORTHOPHOS XG 4 Adjustment...
Page 269: ...ORTHOPHOS XG 5 Service routines...
Page 433: ...ORTHOPHOS XG 6 Repair...
Page 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Page 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Page 531: ...ORTHOPHOS XG 7 Maintenance...
Page 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...