7.7 Checking the tube data
Tabs 7
7.7.3
Fan and temperature
Check whether the fan is functioning
z
Check the function of the fan using service routine S005.4
(see Page 5-31)
If the fan is defective:
‡
Replace the fan (see Page 6-42)
Check whether the temperature sensor
is supplying plausible values
z
Read the temperature in the single tank via service routine S005.5
(see page 5-33)
If the displayed temperature reading is not plausible:
‡
Replace the tube assembly (see Page 6-38)