7 – 29
Tab 7
7.8 Checking the diaphragm
7.
8
7.8.6
Quickshot test exposure
(XG
Plus
and full version only)
Menu: Test exposure
‡
Ceph - Quickshot
CAUTION
It is essential that the PAN needle phantom is removed from bite block holder
of the panoramic X-ray unit before a Ceph exposure is taken; otherwise the
phantom may collide with the sensor.
Opening
CEPH
-
FIXED
POINT
OF
ROTATION
1.
Select the
CEPH
-
FIXED
POINT
OF
ROTATION
submenu.
Opening
QUICKSHOT
2.
Activate the
QUICKSHOT
check box.