7.13 Checking the protective ground wires
7.13
7.13
Checking the protective ground wires
DANGER
Shock hazard! It is essential that you switch the X-ray unit OFF before
replacing any components!
z
Switch
OFF
the line voltage at the main switch of the building installation.
z
Disconnect the power cable and the second protective ground wire from
the building installation.
z
Remove the following cover parts (see section 1.14):
– Profile cover (10)
– Tube assembly cover, front (65)
– Tube assembly cover, rear (70)
Measuring setup for protective ground wire test
V
A
~
A
B
C
D
E
Ammeter
Power source
Measuring point
Central ground wire
Measuring points B-E
Voltmeter