13 Electromagnetic compatibility
Operating Instructions PerioScan
61 29 626 D 3496
D 3496.201.02.05.02
37
båÖäáëÜ
13
Electromagnetic compatibility
13.1
Electromagnetic emission
The
UNIT
is intended for operation in the electromag-
netic environment specified below.
The customer or user of the
UNIT
should make sure that
it is used in such an environment.
The unit meets the requirements for electromagnetic
compatibility (EMC) in compliance with IEC 60601-1-2:
2001.
Emission measurement
Conformity
Electromagnetic environment guidelines
RF emissions according to
CISPR 11
Group 1
The
UNIT
uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions according to
CISPR 11
Class B
The
UNIT
is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
dential purposes.
Harmonics
according to
IEC 61000-3-2
not applicable
Power output
< 50 W
Voltage fluctuations/Flicker
according to
IEC 61000-3-3
not applicable
no significant
flicker