63 22 668 D3561
14
D3561
.
021.01.06
.
02 12.2014
3 On-site installation
Sirona Dental Systems GmbH
3.3 Media quality
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
● Observe EN 1717 for the protection of public drinking water.
Treatment center with disinfection system:
The treatment center fulfills the requirements of EN 1717 and that of
the DVGW (the German Technical and Scientific Association for Gas
and Water). It is intrinsically safe in accordance with worksheet
W540. It can be connected directly to the public drinking water
supply.
Treatment center without disinfection system:
If compliance with EN 1717 is stipulated, appropriate equipment must
be installed beyond the treatment center to protect the public drinking
water system.
Air quality
Air quality
The compressed air for driving the highspeed handpiece, for cooling the
burr drives and for the cooling spray must be free from oil, dry and
hygienically faultless.
Install a steam trap K.
● Permitted air pressure: 5.5bar (80psi) to 7.5bar (109psi)
● Permitted minimum flow volume: 50l/min
Suction pipe
With a vacuum of p
u
> 0.18 bar without flow, the treatment center must
be retrofitted with the "Vacuum limiter" retrofit kit (REF 59 68 826).
● Minimum suction power: 500l/min
● Maximum vacuum: 0.18bar (2.6psi)
Summary of Contents for Sinius
Page 1: ...f o NOKOMNQ pfkfrp L pfkfrp p L pfkfrp qp k W b Installation requirements...
Page 2: ......
Page 57: ......