4 Dimensions, technical data
Sirona Dental Systems GmbH
4.4 Mounting plates
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
4.4
Mounting plates
Adapter plate for Teneo, Sinius
Adapter plate
If the dental treatment center is being installed as a replacement for an
M1 previously installed at this location, an adapter plate is available for
this purpose. You can use the existing drill holes for fastening. Two
additional drill holes are required only in the area of the water unit. The
treatment center is screwed onto the steel plate using M10 screws.
The adapter plate also must be used if the unevenness of the floor
exceeds 2 mm in the vicinity of the base plate.
Note: If a C-Line unit (e.g. C2
+
, M1
+
) was previously installed, the
fastening drill holes must be re-drilled.
Adapter plate thickness: 12 mm
Drilling with floor (7x)
REF 63 07 511
Demo plate for Teneo, Sinius