“LASER STOP” button
Observe all labels on the laser unit.
WARNING
Operation of this laser unit by unauthorized persons must be prohibited in or-
der to prevent incorrect or improper use.
WARNING
Never direct the laser beam or the aiming beam toward the eyes or the thyroid
gland of a person. All persons present in the room (patient, dentist and assis-
tant) must always wear the laser protective goggles delivered along with the
laser unit.
WARNING
Never use optical instruments such as microscopes, eye loupes or magnifiers
together with the original protective goggles. Adequate eye protection is no
longer ensured in this case.
CAUTION
Always cover the optical fiber connection with the special protection caps pro-
vided.
WARNING
The unit is not suitable for use in the presence of anesthetics that are flamma-
ble when in contact with air, oxygen or protoxide.
WARNING
Oxygen-saturated materials such as cotton wool can catch fire due to the high
temperature that the unit reaches during operation. Label removers and flam-
mable solutions used for cleaning and disinfecting the laser unit should be al-
lowed to evaporate before using the device. Be aware of the fire risk caused by
flamable gases.