5 Operating Instructions
båÖäáëÜ
båÖäáëÜ
Here you can select from 10 programs with the arrow keys.
¾
Pressing the "ENTER" key takes you to the selected program
(PP1 for program 1, PP2 for program 2,......)
¾
Confirm the fiber you would like to have for your personal setting
(200µm or 320 µm) with "ENTER" or change it with the "RIGHT ARROW"
key.
After selected the submenu
ª
The green LEDs start flashing, indicating that the laser is in standby mode
now.
ª
The laser is now ready for operation.
ª
If aiming beam is on, please wear protective goggles.
Figure 25: “MANUAL SETTING AREA” submenu
1.
Selected program:
in our example P
P
3:
2.
Laser activation time in milliseconds
: in single pulse or auto-repeat
mode. In the autorepeat mode, the laser unit switches back on after
two seconds
if the foot switch is actuated. If set to any time, laser radiation
will stop after this time, even if you continue to press the foot switch or finger
switch. If set to zero, “------” appears on the display, and the laser remains
activated as long you press the foot switch or finger switch. You can set a
time between 100 ms and 60 s (60000 ms) in 50 ms intervals. When you
press the key for more than 3 s , the value increases at a faster rate.
3.
Laser radiation:
ON when the foot switch or finger switch is actuated.
OFF when the laser is in standby mode.
4.
Selected laser power
: in our example 0.8 W peak power.
Using the
“LEFT ARROW” and “RIGHT ARROW” key, you can adjust the emitted
power between 0.5 W and 7 W in 0.1W increments.
WARNING
Setting the power to excessively high levels increases the risk of overheating
the patient’s soft or hard tissue. Setting the power to excessively low levels
may result in reduced treatment efficacy.
5.
Frequency:
Modulation frequency of the laser. The frequency range is
from 1 to 10,000 Hz. If set to zero, CW (for continuous wave mode) is
displayed. Up to 100 Hz, the parameter value can be increased in 1 Hz
increments. Between 100 Hz and 1,000 Hz, the parameter value can be
increased in 100 Hz increments. Between 1,000 Hz and 10,000 Hz, the
Summary of Contents for sirolaser blue