60 90 570 D 3485
48
D 3485.201.04.09.02
09.2007
9 Technical data
9
Technical data
Laser system
Class IV (according to IEC 60825-1: 2003)
Equipment classification
Class IIb (according to Council Directive 93/42/EEC)
IP degree of protection
Laser unit: IP20; foot switch: IPX5
Wavelength
970 nm +/- 15 nm
Power max.
approx. 7 W CW (optical)
Aiming beam
635 or 650 nm, 1 mW max.
Emission mode
CW (continuous wave) or modulated 1 Hz to 10 kHz
Pulse (chopped mode)
Single or repeated pulse
Pulse duration
100 ms - 60 s in 50 ms intervals
Optical fiber thicknesses
200, 320 and 400
μ
m, NA > = 0.22
Start
Electrical foot switch/finger switch plus electronic access key
Adapter
External, 100 - 240 VAC, 50- 60 Hz
Rating
48 VA
Isolation class
Class 1, type B
Operation parameters
on graphical display
Dimensions
87 × 54 × 190 mm (3.43" × 2.1" × 7.48")
Weight
0.45 kg (1 lb.)
Power supply
The SIROLaser may only be operated with the Sinpro MPU50-105 power
supply provided
NOHD (Nominal Ocular Hazard
Distance)
approx. 1.5 m for 10 s radiation exposure of the unprotected eye
(numerical aperture of the optical fiber NA = 0.22)