60 90 570 D 3485
D 3485.201.04.09.02
09.2007
51
10 Manufacturer’s declaration regarding electromagnetic compatibility
båÖäáëÜ
båÖäáëÜ
RF immunity
The SIROLaser is intended for use in the electromagnetic environment specified below. The end user should ensure that
it is used in such an environment.
Immunity test
EN 60601-1-2
test level
Compliance
level
Electromagnetic environment –
guidance
Conducted RF
EN 61000-4-6
3 Vrms from 150 kHz
to 80 MHz
3 Vrms from 150 kHz
to 80 MHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the unit, including cables, than the
recommended separation distance calcu-
lated from the equation applicable to the fre-
quency of the transmitter.
d = 1.2
√
P from 150 kHz to 80 MHz
d = 1.2
⋅ √
P from 80 MHz to 800 MHz
d = 2.3
⋅ √
P from 800 MHz to 2.5 GHz where
“P” is the maximum output power rating of
the transmitter in watt according to the trans-
mitter manufacturer and “d” is the recom-
mended separation distance in meters.
Radiated RF
EN 61000-4-3
3 Vrms from 80 MHz
to 2.5 GHz
3 Vrms from 80 MHz
to 2.5 GHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less
than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Summary of Contents for sirolaser blue
Page 1: ...k W MVKOMMT b pfoli l f...