5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Instruments and instrument hoses
Operating Instructions TENEO
2.
Touch the
Autopurge function
key.
ª
The
Autopurge
screen is displayed on the touchscreen.
Setting the purge time and starting the autopurge function
The purge time of the instruments may be set between 60 and 180 seconds.
9
The
Autopurge
screen is displayed on the touchscreen.
1.
Use the – and + keys to set the purge time.
2.
Touch the
Start
key.
Error message: Deposit instruments in instrument holders
If the
Deposit instruments
display appears after the autopurge function has
been started, the treatment center has detected that not all of the instruments
have been placed in their holders.
¾
Check the seating of the instruments in the holders marked with a
warning triangle on the touchscreen.
ª
When all of the instruments have been deposited, the autopurge
program continues.
Insert instruments in sanitation bowl
All water carrying instruments and the spray aspirator are inserted in the
sanitation bowl with various adapters. The water flowing over the instruments
is removed from the sanitation bowl by the spray aspirator.
9
The
Insert instruments in sanitation bowl
display appears on the
touchscreen.