5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Vacuum system
Operating Instructions TENEO
Approved care and cleaning agents
Greasing suction handpieces
The disconnection points of the suction handpieces should be greased
following each thermal disinfection or sterilization process and once a week.
Unsuitable greases
1.
Pull the handpiece of the spray aspirator, the saliva ejector and, if
installed, the surgical suction device off of their suction hoses.
2.
Take apart the handpiece of the spray aspirator at its joints.
3.
Regrease the disconnection points and O-rings of the handpieces.
5.4.3
Cleaning and disinfecting the vacuum system
The vacuum system is subjected to septic secretions, saliva and blood daily.
Disinfection at regular intervals is therefore absolutely mandatory for hygienic
reasons.
Approved care and cleaning agents
CAUTION: Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona (see
"Care and cleaning agents" [
).
134°C 273,2°F
3 min
E
D
C
B
A
A
Suction cannula
B
Suction handpiece
C
Handpiece of saliva ejector
D
Surgical suction cannula
E
Intermediate piece
WARNING: Unsuitable greases
Greases that are not food safe may endanger the patient's health Rubber
materials such as e.g. O-rings are corroded by unsuitable greases.
¾
Do not under any circumstances use vaseline or silicone greases.
¾
Use only greases approved by Sirona.
CAUTION: Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona (see
"Care and cleaning agents" [
).