5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Vacuum system
1.
Switch the treatment center off at the main switch.
2.
Pull the filter housings
(B)
off of the connectors on the assistant element.
3.
Disconnect the suction hoses from the filter housings
(B)
.
4.
Take the collectors
(C)
out of the suction hoses. Collect the amalgam
residues in a glass filled with water.
5.
Take off the suction cannulae and rinse out the suction hoses with clean
water.
If the treatment center is equipped with thermodisinfectable suction hoses,
thermal disinfection can be performed after cleaning, see "Thermodisinfecting
the suction hoses" [
.
Assembly is performed in reverse order. Grease the O-rings
(A)
before
reconnecting the suction hoses. For grease, see "Care and cleaning agents".
Make sure that the suction hoses snap into place.
5.4.5
Thermodisinfecting suction hoses
The hoses the treatment center is equipped with as a standard feature cannot
be thermally disinfected. Thermally disinfectable suction hoses are available
as special accessories. They are marked with a turquoise ring, see item
A
in
the drawing below.
Before thermally disinfecting the suction hoses, you must remove them from
the assistant element and clean the filter inserts; see "Cleaning the suction
hoses" [
.
9
The suction hoses are removed and cleaned.
1.
Pull the handpiece
(B)
off of the spray aspirator hose.
2.
Grease the O-rings
(C)
. For greases, see "Care and cleaning agents" [
.
3.
Plug the ends of the handpiece suction hoses onto the hose holders. Use
the adapter
(D)
for the saliva ejector hose.
93°C 197°F
D
C
B
A