5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Components of the water unit
Removing and emptying the sediment container
1.
Disinfect the vacuum system, see "Cleaning and disinfecting the vacuum
system" [
2.
Open the flap
(A)
on the base of the water unit.
3.
Loosen the bayonet catch by turning the sediment container
(B)
counterclockwise.
4.
Pour the excess water out of the sediment container and collect the
amalgam residues. Dispose of these properly together with the amalgam
residues collected from the cuspidor bowl and from the suction hoses;
see "Cleaning the gold trap" [
and "Cleaning and disinfecting the
153]. Authorize a certified waste management
company for this purpose.
Installing the sediment container
1.
Grease the O-ring
(C)
on the sediment container. For greases, see "Care
2.
Hold the sediment container
(B)
so that the latching noses of the bayonet
catch are positioned transverse to the water unit. Rotate the sediment
container
(B)
clockwise.
ª
The sediment container is locked in place.
3.
Close the flap
(A)
.
5.5.7
Cleaning the filter insert of the wet suction
device
The automatic separator and amalgam separator or sediment container are
not installed in the water unit for wet suction. The separation of air and water
and amalgam separation are performed centrally in this case.
In order to ensure that larger solid particles nevertheless cannot enter the
separating unit, the vacuum line in the water unit is equipped with a filter. The
filter insert must be cleaned when the suction power decreases.
A
B
C
B
A