5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Components of the water unit
Operating Instructions TENEO
As long as the treatment center is switched on, water mains in the vacuum
line for technical reasons. Therefore, in order to clean the filter insert this
water must be completely extracted beforehand. Otherwise the remaining
water will flow out when the filter housing is opened. When the treatment
center is switched off at its main switch, the remaining water is automatically
extracted.
Wear gloves
Disinfect the vacuum system
The vacuum system should be disinfected before cleaning the filter insert of
the wet suction device; see "Cleaning and disinfecting the vacuum system" [
.
Opening the chair sub-screen
Opening the sub-screen
9
The
Start program
is displayed on the touchscreen.
¾
Touch the
Sub-screen
key.
ª
The
Start
sub-screen is displayed.
Extracting residual water
¾
Touch the
Extract residual water
key.
ª
As long as the key is flashing orange, water is being extracted from
the water unit. A slurping noise indicates that the water unit is completely
empty. When the key turns gray the extraction process is finished.
Removing and cleaning the filter insert
Catch open during wet suction
9
Water completely extracted from the water unit.
!
NOTE:
Wear gloves when performing the following work.
CAUTION: The flushing and tumbler filling functions can be
activated when the filter housing is open.
Water then escapes from the open filter housing.
¾
Do not switch the flushing and tumbler filling functions on whenever the
filter housing is open.