61 93 556 D3509
D3509
.
201.01.02
.
02
19.09.2008
162
Sirona Dental Systems GmbH
5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Sanitize the treatment center
1.
Open the flap
(A)
on the base of the water unit.
2.
Loosen the bayonet catch of the filter insert
(B)
. Unscrew the filter insert
from the filter housing of the wet suction device counterclockwise.
3.
Properly dispose of the amalgam residues together with the amalgam
residues collected from the cuspidor bowl and from the suction hoses.
Then clean the filter insert under running water in a sink (not in the
cuspidor bowl!).
4.
Grease the O-ring
(C)
of the filter insert. For greases, see "Care and
.
5.
Reinsert the filter insert in the filter housing. Rotate the filter insert
(B)
clockwise.
6.
Close the flap
(A)
.
ª
The treatment center is again ready for operation.
5.6
Sanitize the treatment center
Sanitation effectively combats the reproduction of microorganisms in the
water lines.
Sanitation must be performed:
z
regularly every 4 weeks when the display
days until next sanitation
has
elapsed (1d = days to next sanitation)
z
after longer periods of disuse (> one week)
z
if the germination index exceeds 100 germs per milliliter "Microbiological
water test" [
A
B
C
B
A