5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Sanitize the treatment center
Operating Instructions TENEO
Sanitation comprises several phases and takes at least 24 hours.
Sanitizing additional devices
Preparation
The following preparations should be made prior to beginning sanitation.
1.
Set all burr drives and the SIROSONIC TL scaler to the maximum water
flow rate.
2.
Place all instruments and suction hoses in the instrument holders.
Opening the sanitation screen via touchscreen
9
The
Start program
is displayed on the touchscreen.
1.
Touch the
Sub-screen
key.
ª
The
Start
sub-screen is displayed.
2.
Touch the
San
key.
ª
The
Sanitation
screen is displayed on the touchscreen.
Symbols 1 to 6 stand for the individual sanitation phases as described in the
following. The current sanitation phase is highlighted by an orange rectangle.
CAUTION: Additional devices connected to the external device
connection must not be sanitized with the treatment center.
Additional devices can be damaged. Disinfectant residues can remain in the
additional devices.
¾
Disconnect any additional devices from the treatment center prior to
sanitation.
max.