5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Sanitize the treatment center
7.
Insert the SPRAYVIT valve body of the dentist element
(G)
in the right
adapter and the SPRAYVIT valve body of the assistant element
(H)
in the
left adapter as far as they will go (with valve lever pointing up and locking
button pointing down).
8.
Insert the couplings of the treatment instruments as far as they will go or
until they snap into the adapter.
• white = turbine
I
• green = SL and BL motors
J
• blue = BL ISO motor
K
• red = ultrasonic hose
L
9.
Attach the suction hose
N
and saliva ejector hose
O
(and the surgical
suction device
M
if installed) to the cleaning adapter laterally.
ª
The sanitation bowl is completely fitted.
10.
Touch the
Start
key on the touchscreen.
Sanitation phase 2 – Checking the water flow
First the treatment center checks whether there is water flow through the
instruments.
9
Sanitation phase 2 is highlighted on the touchscreen.
¾
Wait briefly until the water flow has been checked.
ª
If sufficient water flow is present, the treatment center continues with
sanitation phase 3.
Error message: No water flow
If the treatment center detects no water flow through an instrument or through
the tumbler filler, you can try to restore water flow at the location concerned.
If this is not possible, the instrument concerned can be excluded from the
sanitation process.
1.
Check the water flow through the instruments in the holders marked with a
warning triangle on the touchscreen. Set to maximum water flow on the
instruments. Leave all instruments plugged into their adapters.
ª
If the treatment center detects the water flow, the warning triangle will
disappear. If there is sufficient water flow through all instruments,
sanitation phase 3 automatically continues.
2.
If you want to exclude the instruments concerned from sanitation, touch
the
Exclude instrument
key.
ª
The treatment center then continues with sanitation phase 3. The
water lines you excluded are not included in the sanitation process.
NOTE: Pinching of the instrument hoses
Be careful not to pinch the instrument hoses when inserting the instruments.
If the hoses are pinched, the water flow will be obstructed during purging.