5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Sanitize the treatment center
Operating Instructions TENEO
Sanitation phase 3 – Treating water lines with disinfectant
The water is pumped out of the water tank of the water unit via the tumbler
filler. Then the water tank is automatically filled with DENTOSEPT P. This
activates flushing of the instrument hoses, the SPRAYVIT hoses and the
tumbler filler with the disinfectant.
Sanitation phase 4 – Allow 24 hours reaction time
In order to effectively combat germs, you must let the disinfectant react for at
least 24 hours and not longer than 3 days (maximum sanitation time).
9
Sanitation phase 4 is highlighted on the touchscreen.
9
The treatment center displays the remaining reaction time on the
touchscreen next to the sanitation phase 4 field, starting from 24 hours.
9
The treatment center has automatically switched to the Standby mode
1.
Leave all instruments plugged into their adapters.
2.
Make sure that the treatment center is switched off for at least 24 hours,
but not longer than 3 days (maximum sanitation time). If necessary, you
can also switch off the power switch on the base of the treatment center
patient chair.
3.
Switch the treatment center on again after 24 hours and before 3 days
have elapsed.
Error message: Insert instruments in sanitation bowl
If an instrument is accidentally deposited in its holder during the reaction time,
the message
Insert instruments in sanitation bowl
is displayed on the
touchscreen after the treatment center is switched on. The sanitation program
is then completed anyway.
¾
Remove the accidentally removed instrument from its holder and place it
back in the sanitation bowl so that it can be rinsed in sanitation phase 5.
ON
OFF
A
Max.6A
FuseT6.3AH250~
WA
TER AIR
NOTE: Switched-off water and air supplies
If the treatment center is switched on again after 24 hours, the sanitation
process automatically continues with sanitation phase 5. However, if the
water and air supplies are switched off, the disinfectant cannot be rinsed out
of the water lines.
If you switch the treatment center back on after 24 hours have expired, first
check to make sure that the water and air supply lines are open.