7 Malfunctions
Operating Instructions TENEO
Error messages
7
Malfunctions
7.1
Error messages
Any error states of the treatment center that cannot be immediately
recognized by the user and must nevertheless be corrected are displayed on
the touchscreen. If an error state exists, this is indicated by a warning triangle
(A)
in the status line of the touchscreen.
¾
Touch the warning triangle
A
in the footer of the touchscreen (for more
than 2 s).
ª
The error code(s) is(/are) displayed.
The error codes have the following meaning:
Code
Error
Description
Actions required
10
Flow rate of SPRAYVIT on
dentist element too low
The water flow rate is below the limit
value. The heater of the SPRAYVIT is
no longer heated to prevent scalding by
overheated water.
Clean the outlet nozzle of the Sprayvit,
see the "SPRAYVIT L Operating
Instructions" and perform the flow test,
see "Checking the flow rate on the
SPRAYVIT L multifunctional syringe"
[
11
Flow rate of SPRAYVIT on
dentist element too low
12
Tumbler temperature fuse
defective
The temperature fuse was blown by
overheating. The water in the tumbler
filling unit is no longer heated.
Inform your service engineer.
13
Battery of wireless foot control
empty
Operation of the treatment center no
longer possible due to fully depleted
battery.
See "Changing the battery of the
wireless foot control" [
.
14
Disinfectant consumption too
low
The disinfectant tank was not emptied
since its last filling even though many
disinfection cycles were performed.
Inform your service engineer.
15
Water feed too low
The maximum filling time for the mixing
tank is exceeded.
Change the water filter, see "Changing
the water and air filters" [
Check the water pressure of the
drinking water network.
16
Hydrocolloid flow rate
It was detected via the water pump that
the flow rate of the hydrocolloid is too
low.
Check the hydrocolloid hoses for
clogging.
17
Emergency pump operation
Malfunctioning of pump sensors in the
water unit.
Inform your service engineer.
18
Chair movement outside of the
permissible movement limits
or chair drive position not
plausible
If the chair is located outside of the
permissible movement limits or the
position is not plausible, the chair stops
moving.
Inform your service engineer.
19
Injection valve for disinfectant
defective
No disinfectant is added if the injection
valve is defective.
Inform your service engineer.