4 Operation
Sirona Dental Systems GmbH
Patient chair
Operating Instructions TENEO
This can be used to set the cuspidor bowl so that it automatically moves
inward when the mouth rinsing position (S) chair program is selected;
see"Linking the movement of the cuspidor bowl to the mouth rinsing position"
[
.
Via touchscreen
9
The
Start program
is displayed on the touchscreen.
¾
Touch the
S
key briefly (< 2 s).
Via foot control
9
All instruments are in place.
¾
Press the left button of the foot control.
Via the assistant element
¾
Press the
S
key on the assistant element briefly (< 2 s).
4.4.8.3
Using the last position memory function
The last chair position is stored before the patient chair moves to mouth
rinsing position S. When mouth rinsing position key
S
is pressed again, the
treatment center returns to the previously set treatment position.
9
The patient chair is located in any treatment position.
1.
Press the
S
key on the touchscreen, or press the
S
key on the user
interface of the assistant element, or press the left button of the foot
control (with all instruments in place in their holders).
ª
The treatment center moves to the mouth rinsing position.
2.
Press the
S
key on the touchscreen, or on the user interface of the
assistant element, or press the left button of the foot control again.
ª
The treatment center automatically returns to the position where the
patient chair was located prior to the mouth rinsing position.
4.4.8.4
Activating other chair programs
The number of chair programs can be extended to 4 or limited to 2; see
"Showing/hiding chair programs 3 and 4" [
.
Via touchscreen
9
The
Start program
is displayed on the touchscreen.
¾
Touch key
1
,
2
or, if required,
3
,
4
briefly (< 2 s).
S
0
S