4 Operation
Sirona Dental Systems GmbH
Patient chair
Operating Instructions TENEO
4.4.10
Creating chair and shock positioning programs
Chair programs
The six factory set chair programs:
z
Mouth rinsing position S
z
Entry/exit position 0
z
1 and 2
z
3 and 4 if selected
can be individually reprogrammed for each of the six user profiles (A to F).
You can determine whether chair programs 3 and 4 will be displayed in the
chair screen; see "Showing/hiding chair programs 3 and 4" [
9
The
Start program
is displayed on the touchscreen.
1.
Move the patient chair to the required treatment position; see "Moving the
chair manually" [
.
2.
If a motor-driven headrest is installed: Tilt the headrest to the required
treatment position; see "Inclining the headrest" [
3.
Switch the operating light on or off (to program the desired state), see
"Switching the SIROLUX FANTASTIC on/off".
4.
Move the dentist element to the required position; see "Moving the dentist
element" [
5.
Press and hold the desired program key (S, 0, 1, 2 or, if applicable, 3, 4)
(> 2 s).
ª
An audible signal sounds. Your settings are now stored under the
desired program key.
Shock positioning
When the
Shock positio
ning
key is pressed, the patient chair immediately
moves to a position suitable for shock positioning of the patient.
The shock positioning position preset at the factory can be reprogrammed.
1.
Move the patient chair to the desired position.
2.
Press and hold the
Shock positioning
key (> 2 s).
4.4.11
Setting the Massage/Active lumbar support
functions
The patient chair has a gentle massage function and/or an active lumbar
support function.
NOTE: Dentist element movement
You can select the chair programs during which the dentist element is to
move; see "Linking dentist element movement to chair programs" [
.
NOTE: Programming on the assistant element
Chair programs S, 0, 1, 2 can also be programmed on the assistant element
side.