4 Operation
Operating Instructions TENEO
Dentist element
4.5.4.9
Clean
Pressing this key deactivates the complete user interface of the dentist
element. Pressing it again > 3 s reactivates the control panel.
This is used to make sure that no unwanted functions can be accidentally
triggered while cleaning the surface, see "Disinfecting the EasyTouch" [
4.5.4.10
Setup
Used for individual configuration of the treatment center by the user and for
reading out messages by the service engineer, see "Configuration of the
treatment center (Setup)" [
.
4.5.5
Quick setting keys and function levels
Control options
The treatment center offers you two selectable control options in certain
Instrument programs. The instruments can be set either with the quick setting
keys or via the function levels. The selected setting applies to each user
profile, A to F, for the following Instrument programs:
z
z
z
SIROTOM HF electrosurgical handpiece [
Difference between quick setting keys
and function levels
If you use the
Quick setting keys
, keys with five predefined motor speed
values (0.09...40 x1000 rpm) or intensity values (1...100 %) are displayed.
The values shown on the quick setting keys can be selected by touching them
briefly (< 0.5 s).
Intermediate values can be set as follows: If you press and hold a quick
setting key (> 0.5 s) whose value is greater than or equal to the speed or
intensity value displayed in the first line, the value increases. If you press and
hold a quick setting key (> 0.5 s) whose value is less than the speed or
intensity value displayed in the first line, the value decreases. The quick
setting keys are shaded gray for intermediate values.
Motor screen (speed) and ultrasonic screen (intensity) with quick setting keys
When using
function levels
, you have two "storage locations" (E1/E2) at
your disposal for saving settings or recalling them at the push of a button.
These settings can nevertheless be changed during treatment.
A distinction is made between coarse and fine adjustment of the rpm and
intensity settings. If the – or + key is touched briefly (< 0.5 s), the increments
correspond to the quick setting keys (speed: 0.09 or 0.2, 10, 20, ...; intensity:
1, 25, 50, ...). If the – or + key is held (> 0.5 s), intermediate values can also
be set.