61 93 556 D3509
77
D3509
.
201.01.02
.
02
19.09.2008
4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
SIROSONIC TL laser radiation
The
Laser Radiation
warning symbol is displayed in the upper left-hand
corner of the touchscreen in the
Ultrasonic
screen.
The instrument light is generated by a powerful LED located at the instrument
end of the hose. The light is fed through the instrument via a light guide.
4.5.12.2
Setting the intensity
Selecting the intensity with the quick setting keys
9
The SIROSONIC TL scaler is removed from its holder.
9
The
Ultrasonic screen
is displayed on the touchscreen.
¾
Touch one of the quick setting keys in the bottom line key briefly (< 0.5 s).
ª
The quick setting key is highlighted orange. The selected intensity is
displayed in percent in the first line.
Setting intermediate intensity values with quick setting keys
9
The SIROSONIC TL scaler is removed from its holder.
9
The
Ultrasonic screen
is displayed on the touchscreen.
¾
Increasing the intensity: Press and hold a quick setting key whose
intensity value is greater than or equal to the value displayed in the first
line (> 0.5 s).
Reducing the intensity: Press and hold a quick setting key whose
intensity value is less than the value displayed in the first line (> 0.5 s).
ª
The selected intensity is displayed in percent in the first line. The
intensity changes in increments of 1. The quick setting keys are shaded
gray for intermediate values.
WARNING: The ultrasonic hose contains an LED that falls under
laser class 2.
If the ultrasonic handpiece is switched on with the instrument hose removed,
there is a risk of eye damage.
¾
Do not switch the scaler on when the ultrasonic handpiece is removed
from the instrument hose.
¾
Do not stare into the LED at the end of the hose.