61 93 556 D3509
87
D3509
.
201.01.02
.
02
19.09.2008
4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
¾
The
Contra-angle handpie
ce
key is used to set the reduction ratio. The
only available reduction rate at the moment is 20:1.
Contra-angle handpiece for implantology
4.5.14.2.2
Calibrating the burr drive
A calibration must be performed at the start of treatment as well as each time
you change or lubricate the contra-angle handpiece.
The contra-angle handpiece is automatically checked during calibration. This
includes a measurement of motor current at different speeds to assess the
properties of the system.
9
The
Implantology program
is displayed on the touchscreen.
1.
Attach the contra-angle handpiece that you would like to use for the
implantology treatment to the electric motor.
2.
Insert the tool in the contra-angle handpiece. This ensures that the tool is
taken into account in the measurement.
3.
Touch the
Cal
key on the touchscreen.
ª
The button flashes orange.
4.
Hold down the foot pedal throughout the duration of the calibration.
ª
The
Cal
key continues to flash. If the burr drive is calibrated, the key
permanently remains highlighted orange. The calibration is then
completed.
4.5.14.2.3
Setting the speed and torque
In the Implantology function, the speed and torque values of the contra-angle
handpiece, and not those of the motor, are specified. The control electronics
of the burr drive calculate the motor control based on the specified gear
reduction and the speed and torque settings.
9
The
Implantology program
is displayed on the touchscreen.
¾
Set the speed and torque of the contra-angle handpiece using the – and
+ keys.
< 0.5 s coarse adjustment, > 0.5 s fine adjustment
ª
The selected speed is displayed in the first line in rpm (revolutions per
minute). The torque is displayed in the second line in Ncm (Newton
centimeter).
WARNING: Only the Sirona contra-angle handpiece IMPLANT 20:1
may be used for the Implantology function.
Instruments from other manufacturers can lead to malfunctions. Outside
instruments may be improperly calibrated for implantology.
¾
Use only the IMPLANT 20:1 contra-angle handpiece from Sirona for
implantology.
¾
Check whether the reduction ratio on the touchscreen agrees with the
value specified on the contra-angle handpiece being used.
NOTE:
Use only Sirona instruments to guarantee correct calibration.