4 Operation
Operating Instructions TENEO
Dentist element
9
The
Simple Endodontics program
(shown here) or the
Advanced
Endodontics program
is displayed on the touchscreen. The
Advanced
Endodontics program
can be opened by touching the
Sub-screen
key
(two cascaded rectangles).
1.
Attach the contra-angle handpiece that you would like to use for the
treatment to the electric motor.
2.
Insert a file in the contra-angle handpiece. This ensures that the file is
taken into account in the measurement.
3.
Touch the
Cal
key on the touchscreen.
or
¾
Press the right button of the foot control.
ª
The button flashes orange.
4.
Hold down the foot pedal throughout the duration of the calibration.
ª
The
Cal
key continues to flash. If the burr drive is calibrated, the key
permanently remains highlighted orange. The calibration is then
completed.
4.5.14.3.4
Setting the speed and torque
You can modify the parameters of the file if you don't want to use the default
file settings.
In the Endodontics function, the speed and torque values of the contra-angle
handpiece, and not those of the motor, are specified. The control electronics
of the burr drive calculate the motor control based on the specified gear
reduction and the speed and torque settings.
9
The
Advanced Endodontics program
is displayed on the touchscreen.
9
A file is selected.
¾
Set the speed and torque of the contra-angle handpiece using the – and
+ keys.
< 0.5 s coarse adjustment, > 0.5 s fine adjustment
ª
The selected speed is displayed in the first line in rpm (revolutions per
minute). The torque is displayed in the second line in Ncm (Newton
centimeter).
For endodontics, the maximum values recommended by the file manufacturer
are displayed above the speed and torque values in fine gray print. They
serve as an orientation when setting.
Incorrect speed and torque settings
NOTE: Torque adjustment
The maximum adjustable torque depends on the system motor and the speed
settings.
WARNING: Improperly selected speeds and torque values
endanger the patient.
Treatment errors, e.g. breaking of a file, may result from incorrect settings.
¾
Observe the manufacturer's instructions regarding file systems.