61 93 556 D3509
D3509
.
201.01.02
.
02
19.09.2008
102
Sirona Dental Systems GmbH
4 Operation
Operating Instructions TENEO
Assistant element
4.6.4.5
Chair programs
The following chair programs can be selected and programmed on the
assistant element:
z
Mouth rinsing position (S)
z
Entry/exit position (0)
z
Chair programs 1 and 2
For details, see "Moving the patient chair via chair programs" [
Chair programs can also be programmed from the assistant element side;
see "Programming chair programs and shock positioning" [
4.6.4.6
Headrest
The treatment chair is adjusted to the patient's stature by moving the headrest
in or out.
4.6.4.7
Composite function
Description of composite function
With the composite function, the operating light can be operated at a reduced
brightness of < 8,000 lux.
This function is required to prevent premature curing of composite fillings.
¾
Press the
Composite Function
fixed key.
ª
If the composite function is switched on, the LED on the fixed key
lights up on the assistant and dentist elements. The light intensity of the
operating light is > 8,000 lux.
4.6.4.8
Operating light
The light intensity of the operating light is > 24,000 lux.
The operating light is always switched on at the set brightness level.
¾
Press the
Operating Light
fixed key.
ª
If the operating light is switched on, the LED on the fixed key lights up
on the dentist and assistant elements.
4.6.5
Suction handpieces
The assistant element can be equipped with a maximum of three suction
handpieces.
Cannula attaches itself to oral mucosa
WARNING: The cannula attaches itself to the oral mucosa.
The patient's oral mucosa is irritated by the vacuum.
¾
Make sure to hold the suction cannula in such a way that the cannula
opening cannot attach itself to the oral mucous membranes by accident.
Sirona recommends using spray aspirator cannulae with ventilation
holes "see Spare parts, Consumables [