4 Operation
Operating Instructions TENEO
Water unit
Filling the tumbler
9
The tumbler must not be transparent.
9
The tumbler must always be placed in the center of the area provided for
it.
¾
Place the tumbler under the tumbler filler.
ª
The tumbler is filled automatically.
ª
After the preset filling level has been reached, the water flow stops
automatically.
If required, the tumbler can be refilled manually by pressing the
corresponding
Tumbler filling
key; see "Setting the tumbler filling function" [
.
Setting the filling level
The tumbler filling level can be preset.
1.
Open the maintenance flap
(A)
on the base of the water unit.
2.
Set the filling level with the potentiometer
B
.
4.7.3
Adjusting the water amount for flushing
The water amount and water pressure for flushing can be adjusted with a
valve.
1.
Open the maintenance flap
(A)
on the base of the water unit.
2.
Adjust the water amount with the valve
(B)
.
NOTE: Tumbler filling after switch-on
If an empty tumbler is standing below the tumbler filler when the treatment
center is switched on with the main switch, the tumbler will not be filled
automatically. To activate automatic tumbler filling, remove the tumbler briefly
and then put it back again.