4 Operation
Operating Instructions TENEO
Configuration of the treatment center (setup)
4.13.5.1
Preselecting quick setting keys with SaveMode/
with DropMode or function levels
Settings can be made in the instrument programs using either the quick
adjustment keys (1...100) or two freely programmable function levels (E1,
E2). For more information, please refer to "Quick setting keys and function
levels" [
.
When using the quick adjustment keys, you can choose one of two options for
saving the settings you made in the instrument program:
z
SaveMode – The
Memory
key is displayed in the instrument programs:
The settings made in the instrument program will be saved after the
instrument is placed in its holder only if the Memory key was pressed and
held beforehand (> 2 s).
z
DropMode –
Memory
key hidden in Instrument programs:
When the instrument is deposited, the settings made in the Instrument
program will automatically be saved.
One of the following presettings can be selected:
z
Field 1: Quick setting keys with SaveMode
z
Field 2: Quick setting keys with DropMode
z
Field 3: Function levels
¾
Touch the
Select memory type
key.
ª
The selected field is highlighted orange.
4.13.5.2
Switching the motor aftercooling on/off
After the burr drive is deposited in its holder, the motor can be automatically
aftercooled.
¾
Touch the
Motor aftercooling
key.
ª
If the key is highlighted orange, the motor aftercooling is activated.
4.13.5.3
Switching afterblow on/off
After an instrument is deposited in its holder, the cooling spray remaining in
the instrument head or in the tip of the instrument is automatically blown out
by a brief activation of the chip blower (air puff).
¾
Touch the
Afterblow
key.
ª
If the key is highlighted orange, the afterblow function is activated.
4.13.5.4
Setting the spray temperature
The spray temperature of all instruments on the dentist element except for the
SPRAYVIT multifunctional syringe can be adjusted.
The spray temperature of the SPRAYVIT L multifunctional syringe can be
adjusted separately, see "Setting the media temperature and instrument light"
[
.
¾
Use the – and + keys to set the spray temperature.