5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Basics
Operating Instructions TENEO
Change the water and air filters [
Sanitize the treatment center [
or
empty the sediment container [
clean the filter insert of the wet suction device [
For a quick overview of the required work, see "Care and cleaning plan for the
TENEO
®
treatment center".
5.1.2
Care and cleaning agents
Care and cleaning agents
You will find a list of all agents approved at the time of delivery in the
documentation folder supplied with your treatment center.
A continuously updated list of approved agents can be downloaded from the
Internet at
"www.sirona.com"
. In the navigation bar, go to the menu items
"SERVICE"
/
"Downloads"
and then open the document
"Care and
cleaning agents"
.
If you do not have any access to the Internet, you can order the list in one of
the following two ways:
z
Order from your local dental depot
z
Order from Sirona:
Tel: ++49 (0) 62 51 / 16-16 16
Fax: ++49 (0) 62 51 / 16-18 18
Order No.:
59 70 905
5.1.3
Microbiological water test
Perform the microbiological water test at regular intervals and after longer
periods of disuse > 1 week; see also "Media quality" [
. Start the
checkups in intervals of no more than two weeks and adjust the time intervals
depending on the results. In addition to running laboratory tests, you also can
use the "total count tester" as a simple means of performing this test.
The first three testers will be supplied free of charge if you return the attached
voucher. Please state your address and the serial number of the treatment
center, see "Installation Report / Warranty Passport", or specify the serial
number on the rating plate affixed to the base of the chair.
To reorder the motor total count tester, see "Spare parts and consumables"
[
.
The cardboard disk contains a dehydrated nutrient medium. It is activated by
the sample and serves as culture medium for a number of bacteria. The
number of germs provides information on the hygienic condition of the water.
CAUTION: Approved Care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona!
NOTE: Shelf life of the total count tester
The maximum shelf life of the total count tester is 1 year after the date of
receipt.