5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Surfaces
¾
Press one side of the head pad
(A)
inward from the outside (to unlock it)
and remove it.
The head pad can be snapped back into the headrest by applying slight lateral
pressure.
Scratched surfaces
To reorder the head support, see "Spare parts and consumables" [
5.2.7
Thermally disinfect the instrument holder of the
dentist element
The instrument holder can be removed for easier cleaning or thermal
disinfection.
Removing the instrument holder
1.
Remove all instruments from the holder.
2.
Swivel both of the levers
(A)
below the instrument holder to the rear.
ª
The latch of the instrument holder is released.
3.
Grasp underneath the instrument holder and raise it at the rear.
ª
The instrument holder tilts to the front slightly and can be removed
from the dentist element in an upward direction.
The removable silicone mat on the dentist element can be sterilized.
Inserting the instrument holder
9
The levers
(A)
are swiveled to the rear.
1.
Fit the front recesses
(B)
of the instrument holder onto the bolts in the
dentist element.
2.
Gently press the instrument holder into the dentist element.
3.
Hold the instrument holder firmly and swivel both levers
(A)
toward the
front.
ª
The instrument holder is locked into the dentist element.
Ball stopper
If a ball stopper is inserted in an unused instrument holder, it can be pushed
out of the holder from the rear to facilitate cleaning or thermal disinfection.
WARNING: Deep scratches or damage hinder the cleaning and
disinfection of surfaces.
The health of patients may thus be endangered.
¾
Replace the head support if it is heavily scratched or damaged.