5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Instruments and instrument hoses
Operating Instructions TENEO
Motor holder
A sterilizable motor holder is available for surgical procedures; see also
"Preparing the treatment center for sterile operation" [
5.2.8
Thermally disinfect the instrument holder of the
assistant element
The instrument holders can be removed for easier cleaning or thermal
disinfection.
Removing the instrument holder
1.
Remove all instruments from the instrument holder.
2.
Pull the latch
(A)
underneath the assistant element.
ª
The latch of the instrument holder is released.
3.
Remove the instrument holder.
The removable silicone mat on the assistant element can be sterilized.
Inserting the instrument holder
¾
First insert the instrument holder in the groove underneath the user
interface. Then push the holder forward and downward until it locks in
place.
ª
The instrument holder is locked into the dentist element.
5.2.9
Clean the foot control
Cleaning the foot control regularly improves its stability.
¾
Clean the bottom plate of the foot control with a moist cloth (water).
5.3
Instruments and instrument hoses
5.3.1
Rinsing water lines
Microorganisms can grow in the water lines of the treatment center. Use a
large amount of water for rinsing the lines prior to starting patient
appointments.
134°C 273,2°F
3 min
H O
2