5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Instruments and instrument hoses
¾
Activate the cuspidor flushing of the cuspidor bowl for at least one minute.
5.3.2
Purge water lines (purge function)
To reduce the amount of germs, the water lines of the water carrying
instruments of the dentist element and the SPRAYVIT multifunctional syringe
of the assistant element are purged with water.
For the purge function, individual instruments are removed from their holders
and held over the cuspidor bowl for purging. The water lines of all removed
instruments are then purged simultaneously.
It is also possible to purge the water lines automatically, see "Purging the
water lines automatically (autopurge function)" [
.
Preparation
The following preparations should be made before you begin to purge the
water lines.
1.
Set all of the instruments to be purged to their maximum water flow rate.
2.
Place all instruments in their instrument holders.
Opening the Purge screen
9
The
Start program
is displayed on the touchscreen.
1.
Touch the
Sub-screen
key.
ª
The
Sub-screen
is displayed.
2.
Touch the
Purge function
key.
ª
The
Purge
screen is displayed on the touchscreen.
max.