5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Instruments and instrument hoses
Canceling the purge function
In case of an error message, deposit the instruments in their holders or,
during purging, the purge function can be canceled.
¾
Touch the
Stop
key on the touchscreen.
5.3.3
Automatically purge water lines (autopurge
function)
The autopurge function enables automatic purging of all water carrying
instruments in the dentist element, of the SPRAYVIT multifunctional syringe
in the dentist and assistant elements and of the tumbler filling unit.
All of the instruments inserted in the sanitation bowl will be purged when the
autopurge function is activated. If the instruments remain in the sanitation
bowl after the treatment center is switched off, the purging process will
automatically be started again the next time the treatment center is switched
on.
Execute the autopurge function:
z
before starting work
z
at the end of the work day
It is also possible to purge individual instruments; see "Purging water lines
(purge function)" [
Preparation
The following preparations must be made before you start purging the water
lines.
1.
Set all burr drives and the SIROSONIC TL scaler to the maximum water
flow rate.
2.
Place all instruments in their instrument holders.
Opening the Autopurge screen
9
The
Start program
is displayed on the touchscreen.
1.
Touch the
Sub-screen
key.
ª
The sub-screen is displayed.
max.