5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Instruments and instrument hoses
Operating Instructions TENEO
5.3.4.2
Disinfecting/sterilizing the Mini L.E.D. curing light
Approved care and cleaning agents
1.
Remove the connecting cable of the Mini L.E.D. by rotating the handpiece.
2.
Pull out the light guide and remove the glare shield.
3.
Sterilize the light guide and the glare shield at 135° C, 2 bar for 3 min.
4.
Disinfect the handpiece of the Mini L.E.D.
5.
Screw the sterilized light guide and glare shield back onto the Mini L.E.D.
6.
Reconnect the handpiece of the Mini L.E.D. to the connecting cable.
The following points should also be observed when operating the Mini L.E.D.:
z
Always use the glare shield to protect your eyes.
z
Check the light guide after each use. Make sure that the light guide is in
perfect condition.
z
No composite material residues may be present on the light guide.
Immediately remove any residue.
z
If you find any damage, replace the light guide, since damage will impair
its performance considerably.
5.3.4.3
Cleaning/disinfecting the SiroCam intraoral digital
camera
Due to its shape, the SiroCam digital intraoral camera takes hygienic
requirements into account and therefore has no spots that are difficult to
reach. It can be wiped with surface disinfectants.
Approved care and cleaning agents
Scratch-sensitive lens window
For intraoral use of the camera, hygienic protective covers should be used,
see "Use of hygienic protective covers".
5.3.4.4
Check the flow rate on the SPRAYVIT L
multifunctional syringe
If the flow rate is less than 135 ml/min with the water button fully actuated,
there is a risk of excessively hot water being emitted.
CAUTION: Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona (see
"Care and cleaning agents" [
).
134°C 273,2°F
3 min
CAUTION: Approved care and cleaning agents
Use only care and cleaning agents which have been approved by Sirona (see
"Care and cleaning agents" [
).
CAUTION: The lens window is sensitive to scratches.
Deep scratches in the lens window impair image quality.
¾
Protect the lens window against scratching. Disinfect it with a soft, lint-
free cloth.