5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Vacuum system
Risk of scalding with SPRAYVIT (cleaning)
Perform the following measurement to exclude a patient risk.
¾
Fill a measuring cup up to the 100 ml mark with the water button fully
actuated while measuring the required filling time.
ª
The filling time must not exceed 45 seconds.
5.3.5
Changing the cotton wool roll on the turbine
hose
Return air containing a small amount of turbine oil is emitted at the unit end of
the turbine hose. This oil is absorbed by the cotton wool roll in the transparent
collecting cup.
1.
Push the cup down and remove the cotton wool roll.
2.
Insert a new cotton wool roll.
3.
Push the cup up again.
5.4
Vacuum system
5.4.1
Purge the vacuum system.
To ensure that the vacuum system is always ready for use, you must aspirate
a large glass of clear, cold water through the suction hoses used after each
patient, especially after each intervention during which the patient loses
blood.
During prolonged treatments, you must aspirate a glass of water at least
every 60 minutes.
5.4.2
Sterilize and disinfect the suction handpieces
Sterilization/disinfection
All parts of suction handpieces can be sterilized and thermally disinfected.
WARNING: If the flow rate is insufficient, hot water may be emitted
by the SPRAYVIT.
The patient could thus be scalded.
¾
Check the water flow rate prior to use.
¾
Clean the nozzle according to the SPRAYVIT Operating Instructions or
have the treatment center checked by a service engineer.
100ml
max.45 sec