5 Care and cleaning instructions for the practice team
Sirona Dental Systems GmbH
Components of the water unit
Operating Instructions TENEO
4.
Plug the hose holders onto the rails in the (Miele) thermodisinfector and
place the suction hoses on the wire basket provided for that purpose.
5.
Thermally disinfect the suction hoses, filter housings and collectors at
max. 93° C.
To reorder the hose holders for the Miele thermodisinfector, see "Spare parts
and consumables" [
.
5.5
Components of the water unit
5.5.1
Clean the gold trap
The gold trap retains larger solid particles to prevent them from being washed
down the drain in the cuspidor bowl. In this way, the amalgam rotor must be
replaced and the sediment container must be emptied less often.
Amalgam residues
1.
Remove the gold trap from the drain of the cuspidor bowl.
2.
Remove the amalgam residues from the gold trap. Amalgam residues
must be disposed of separately.
3.
Clean the gold trap.
4.
Reinsert the gold trap.
5.5.2
Clean and disinfect the cuspidor bowl
The cuspidor bowl, the tumbler holder
(A)
and the tumbler outlet
(B)
can be
sprayed and wiped with surface disinfectants.
Clean and disinfect the cuspidor bowl with a special cleaning agent. This
agent will also care for the drain lines of the cuspidor bowl.
CAUTION: Amalgam residues must not enter the public sewage
system.
Amalgam is a mercury compound that is hazardous to water.
¾
Do not dispose of amalgam residues in a sink.
¾
Collect amalgam residues e.g. in a closed container with water. Dispose
of amalgam residues, e.g. when replacing the amalgam rotor by filling the
amalgam residues into the amalgam rotor or when emptying the
sediment container.