5 Care and cleaning instructions for the practice team
Operating Instructions TENEO
Foot control and connection box
Sanitation phase 5 – Rinsing water lines with water
The disinfectant is rinsed out of the SPRAYVIT hoses, instrument hoses and
tumbler filling unit with water. This takes several minutes.
Sanitation phase 6 – Returning the instruments to their holders
After the water lines have been rinsed, the instruments and suction hoses can
be removed from the sanitation bowl and deposited in their holders again.
9
The
Return instruments to their holders
display appears on the
touchscreen.
1.
Place the SPRAYVIT sleeves, straight and contra-angle handpieces and
suction cannulae back on the instruments and suction hoses. Redeposit
all instruments and suction hoses in the instrument holders.
2.
Remove the sanitation bowl and adapter from the cuspidor bowl.
ª
The sanitation process is finished. The treatment center is again
ready for operation.
Aborting the sanitation process
Sanitation can be canceled during some sanitation phases, e.g. if the
treatment center urgently must be switched back on before the 24 hours have
expired. The
Stop
key shows whether cancellation is possible.
9
The sanitation process must have been started.
¾
Touch the
Stop
key on the touchscreen.
ª
The sanitation procedure is canceled. If the water lines have not yet
been treated with disinfectant (sanitation phase 3), the treatment center
automatically proceeds directly to sanitation phase 6. If the sanitation
process is canceled during the reaction time, the water lines are initially
rinsed with water (sanitation phase 5).
5.7
Foot control and connection box
5.7.1
Changing the battery of the wireless foot
control
The wireless foot control is powered by a battery. An empty battery is
detected and displayed by the system. The battery can be changed by the
user.
For the battery type, see "Spare parts and consumables" [
The housing of the wireless foot control must be opened to change the
battery. Touch a grounded metal part before opening the housing to prevent
damage to the PC board due to electrostatic discharge.
¾
Prior to changing the battery, switch the treatment center off at the main
switch. This prevents accidental triggering of unwanted functions.