4 Operation
Operating Instructions TENEO
Patient chair
If the armrest is removed, the supplied cover must be inserted in the mount
on the patient chair instead of the armrest in order to bridge the safety switch.
If the safety switch is not bypassed with the cover when the armrest is
removed, the treatment center will be disabled by the triggered safety switch,
see also "Safety stop" [
1.
Press and hold the release button on the bottom side of the armrest and
pull out the armrest.
2.
Insert the cover instead of the armrest.
ª
The cover automatically locks in place. The safety switch of the
armrest is bypassed.
Proceed in reverse order when installing the armrest. Press the tab of the
cover with one finger to release it from the side of the chair.
4.4.5
Simple/Advanced Start program operating
mode
Function of the Start programs
In the
Start prog
ram
, you can select functions associated with movement of
the patient chair. Furthermore, you also can access additional general
functions via the
Start p
rog
ram
.
Difference between Simple and Advanced Start program modes
In the
Simple Start program
operating mode, only the chair program function
keys and, if present, the
Move headrest in/out
function are displayed in the
screen.
Simple Start program operating mode
In the
Simple Start program
operating mode, the remaining functions are
listed in the separate
Manual Chair Adjustment
screen. This is achieved via
the additional
Manual Chair Adjustment
key.
CAUTION: A swiveling armrest does not have an end stop that
would prevent a collision when fitted in the left-hand position.
Fitting a swiveling armrest in the left-hand position can lead to collisions with
the tumbler during certain chair movements.
¾
Only use the swiveling armrest on the right-hand side of the treatment
center.