4 Operation
Operating Instructions TENEO
Dentist element
9
The
Sub-screen
of the removed instrument is displayed on the
touchscreen.
1.
Press and hold the
Instrument light
key (> 2 s).
ª
The
Instrument Light
settings screen is displayed. The instrument
light comes on.
2.
Use the – and + keys to adjust the brightness of the instrument light.
3.
Touch the
Return
key.
ª
The
Instrument Light
settings screen is hidden immediately. The
settings are saved. If the
Instrument Light
key is highlighted orange, the
function is activated.
4.5.8.5
Preselecting the instrument light operating voltage
The original Sirona halogen lamp is usually operated at 3.6 V. Voltages over
3.8 V destroy the lamp. The operating voltage can be adjusted for lamps
made by other manufacturers.
Setting the operating voltage
9
The
Turbine, Motor
or
Ultrasonic screen
is displayed on the touchscreen.
1.
Press and hold the
Setup
fixed key (> 2 s).
ª
The Setup program of the corresponding instrument is displayed.
2.
Use the – and + keys to adjust the maximum operating voltage of the
instrument light.
The instrument light voltage of the SPRAYVIT L multifunctional syringe can
also be adjusted, see "Setting the media temperature and instrument light" [
.
4.5.8.6
Setting the foot control as a direct starter or speed
foot control
The foot control can be set as a direct starter or as a speed foot control in the
motor or SIROSONIC TL scaler sub-screen:
z
Direct starter
When the foot control is actuated, the instrument is switched on with the
set speed and intensity.
CAUTION: The operating voltages of different lamps may vary.
Overvoltages can lead to damage.
¾
When changing lamps, make sure that the operating voltage is properly
set for the new lamp.